Removal of UroLume endoprosthesis: experience of the North American Study Group for detrusor-sphincter dyssynergia application. Academic Article uri icon

Overview

abstract

  • PURPOSE: We present the experience of the North American UroLume Multicenter Study Group with removal of the UroLume endoprosthesis. MATERIALS AND METHODS: A total of 160 neurologically impaired patients were enrolled in the North American UroLume Multicenter Study Group for detrusor external sphincter dyssynergia application. Analysis was performed in 2 groups of patients in which the device was removed during insertion and after implantation, respectively. RESULTS: Device retrieval was required during insertion in 21 patients (13%) mainly due to misplacement or migration in 17. Extraction was done with minimal complications and in all but 2 cases subsequent UroLume implantation was successful. Of 158 men with the device in place 31 (19.6%) required removal. In 34 procedures 44 devices were removed, mainly due to migration. Time from implantation to removal ranged from 4 days to 66 months (mean 22 months). The UroLume was removed en bloc in 20 cases and in parts or wire by wire in 19. The majority of patients had no or minimal complications after extraction. Only 2 patients had serious temporary complications, including bleeding and urethral injury, with no lasting consequences. No malignancy developed as a result of UroLume insertion. CONCLUSIONS: While there is a potential for urethral injury and bleeding, UroLume endoprosthesis removal is largely a simple procedure with minimal complications and consequences.

publication date

  • March 1, 2000

Research

keywords

  • Stents
  • Urinary Bladder, Neurogenic

Identity

Scopus Document Identifier

  • 18544406752

Digital Object Identifier (DOI)

  • 10.1016/s0022-5347(05)67801-9

PubMed ID

  • 10687974

Additional Document Info

volume

  • 163

issue

  • 3