Optimization of view ordering for motion artifact suppression. Academic Article uri icon

Overview

abstract

  • Motion artifacts in MRI may be reduced by optimized view ordering. Extensive simulations of view-ordering techniques were performed on high-resolution phantom images to determine the best strategy for distributing motion in k-space. Although not exhaustive, simulation results indicate that minimizing motion at the center of k-space is critical to overall image quality. For 2D imaging, using edge-center-edge view order and setting the readout direction parallel to the direction of the motion produces the sharpest point spread function and the lowest image energy error. For 3D imaging, using an edge-center-edge view order proves to be the optimum choice in general. Given these observations, several important issues regarding the measurement of motion effects are discussed.

publication date

  • September 1, 2001

Research

keywords

  • Image Enhancement
  • Magnetic Resonance Imaging

Identity

Scopus Document Identifier

  • 0034758970

Digital Object Identifier (DOI)

  • 10.1016/s0730-725x(01)00426-x

PubMed ID

  • 11595366

Additional Document Info

volume

  • 19

issue

  • 7