Failure of automatic capture verification in an epicardial pacemaker system.
Overview
abstract
Epicardial pacing is frequently used in pediatric patients with congenital heart disease. Automatic threshold determination has been reported to be safe in epicardial lead systems. We report a case of falsely low ventricular thresholds determined by automatic capture verification in a patient with complete loss of capture due to a lead fracture.