Comparative evaluation of an II based and a flat panel based cardiovascular fluoroscopy system within a clinical environment. Academic Article uri icon

Overview

abstract

  • The image quality and dose parameters from a 2004 Siemens Axiom Artis dBC cardiac biplane with flat panel detector were evaluated and compared to similar parameters evaluated for a 1977 Toshiba DPF 2000A biplane cardiac unit with a conventional image intensifier. Image quality assessment was performed with the Westmead test object; using solid water as a patient equivalent absorber. The patient dose comparison of the two systems is based on dose area product meter readings for 1512 patient cases recorded over 6 months following installation of the Siemens flat panel digital unit. The image quality results indicate that: (a) high contrast resolution was better with the digital flat panel unit, (b) low contrast resolution is similar between systems, and (c) the threshold contrast of the flat panel system is the same or inferior to that of the image intensifier system. Input dose to the surface of the flat panel detector showed a strong dependence on field size, similar to the behaviour of image intensifier system. For the most common clinical procedure--Left Heart Study via Judkins--the average total dose area product reading was 64.0 Gy-cm2 against 67.7 Gy-cm2 for the digital and conventional units respectively (p = 0.27) indicating no significant difference in dose performance between the two x-ray machines.

publication date

  • September 1, 2005

Research

keywords

  • Fluoroscopy
  • Heart
  • Radiographic Image Enhancement
  • Radiographic Image Interpretation, Computer-Assisted
  • X-Ray Intensifying Screens

Identity

Scopus Document Identifier

  • 27244453256

Digital Object Identifier (DOI)

  • 10.1007/BF03178708

PubMed ID

  • 16250468

Additional Document Info

volume

  • 28

issue

  • 3