Automatic algorithm for correcting motion artifacts in time-resolved two-dimensional magnetic resonance angiography using convex projections. Academic Article uri icon

Overview

abstract

  • Time-resolved contrast enhanced magnetic resonance angiography (MRA) may suffer from involuntary patient motion. It is noted that while MR signal change associated with motion is large in magnitude and has smooth phase variation in k-phase, signal change associated with vascular enhancement is small in magnitude and has rapid phase variation in k-space. Based upon this observation, a novel projection onto convex sets (POCS) algorithm is developed as an automatic iterative method to remove motion artifacts. The presented POCS algorithm consists of high-pass phase filtering and convex projections in both k-space and image space. Without input of detailed motion knowledge, motion effects are filtered out, while vasculature information is preserved. The proposed method can be effective for a large class of nonrigid motions, including through-plane motion. The algorithm is stable and converges quickly, usually within five iterations. A double-blind evaluation on a set of clinical MRA cases shows that a completely unsupervised version of the algorithm produces significantly better rank scores (P=0.038) when compared to angiograms produced manually by an experienced radiologist.

publication date

  • March 1, 2006

Research

keywords

  • Algorithms
  • Magnetic Resonance Angiography
  • Motion

Identity

Scopus Document Identifier

  • 33644794919

Digital Object Identifier (DOI)

  • 10.1002/mrm.20806

PubMed ID

  • 16463347

Additional Document Info

volume

  • 55

issue

  • 3