A new obstetrical polyurethane versus stainless steel forceps: a comparison of forces generated to the base of the fetal skull during simulated deliveries. Academic Article uri icon

Overview

abstract

  • BACKGROUND: Decreasing the maximum force applied during traction to the base of the fetal skull using a less rigid polyurethane forceps is the basis of this study. Our hypothesis was that less force would be generated with polyurethane forceps than with steel forceps. OBJECTIVE: To test a new soft polyurethane obstetrical forceps for maximal force generated to the base of the skull during simulated occiput anterior deliveries and to compare this to a similar shaped steel forceps. METHODS: After designing a prototype polyurethane forceps, we used a pelvic manikin model and a fetal manikin model. Force and load sensors were attached at the inner tips of the distal forceps blade. A Tekscan 201 (accurate for measuring 0-25 pounds of force) 0.0008 inches flexible printed circuit was used that measured contact forces. Forceps with an attached calibrated sensor were applied to the fetal head while inside the pelvic model. RESULTS: The median maximum traction force at the base of the fetal skull was 4.60 pounds (range 4.3-4.62) for polyurethane forceps vs. 9.52 pounds (range 9.22-9.52) for steel forceps (P=0.027). CONCLUSION: The polyurethane forceps applied 50% less overall mechanical force than the steel forceps at the tip of the forceps and base of the skull during simulated occiput anterior outlet deliveries.

publication date

  • January 1, 2009

Research

keywords

  • Delivery, Obstetric
  • Fetus
  • Obstetrical Forceps
  • Skull

Identity

Scopus Document Identifier

  • 74949104624

Digital Object Identifier (DOI)

  • 10.1515/JPM.2009.136

PubMed ID

  • 19909222

Additional Document Info

volume

  • 37

issue

  • 6