Transcranial electric motor evoked potential monitoring during spine surgery: is it safe? Academic Article uri icon

Overview

abstract

  • STUDY DESIGN: Retrospective review. OBJECTIVE: To report on the safety of repetitive transcranial electric stimulation (RTES) for eliciting motor-evoked potentials during spine surgery. SUMMARY OF BACKGROUND DATA: Theoretical concerns over the safety of RTES have hindered broader acceptance of transcranial electric motor-evoked potentials (tceMEP), despite successful implementation of spinal cord monitoring with tceMEPs in many large spine centers, as well as their apparent superiority over mixed-nerve somatosensory-evoked potentials (SSEP) for detection of spinal cord injury. METHODS: The records of 18,862 consecutive patients who met inclusion criteria and underwent spine surgery with tceMEP monitoring were reviewed for RTES-related complications. RESULTS: This large retrospective review identified only 26 (0.14%) cases with RTES-related complications; all but one of these were tongue lacerations, most of which were self-limiting. CONCLUSIONS: The results demonstrate that RTES is a highly safe modality for monitoring spinal cord motor tract function intraoperatively.

publication date

  • June 1, 2011

Research

keywords

  • Evoked Potentials, Motor
  • Monitoring, Intraoperative
  • Orthopedic Procedures
  • Spinal Cord Injuries
  • Spine
  • Transcranial Magnetic Stimulation

Identity

Scopus Document Identifier

  • 79957846817

Digital Object Identifier (DOI)

  • 10.1097/BRS.0b013e3181ecbe77

PubMed ID

  • 21217447

Additional Document Info

volume

  • 36

issue

  • 13