Reinforcement of HeartMate II bend relief connection: champagne bottle technique. Academic Article uri icon

Overview

abstract

  • HeartMate II (Thoratec Corp, Pleasanton, CA) continuous-flow ventricular assist devices may be exchanged through a subcostal approach, where the device body can be accessed easily and safely. The existing bend relief needs to be reattached to the new device body at the end of the operation, and this reattachment often results in a malalignment between the outflow graft, the bend relief, and the device body. This causes unexpected detachment and subsequent graft kinking. We present a new, simple technique to reinforce the bend relief connection.

publication date

  • April 1, 2013

Research

keywords

  • Heart Failure
  • Heart-Assist Devices

Identity

Scopus Document Identifier

  • 84875306601

Digital Object Identifier (DOI)

  • 10.1016/j.athoracsur.2012.12.023

PubMed ID

  • 23522236

Additional Document Info

volume

  • 95

issue

  • 4