Disconnection of the rubber tip of arrow-trerotola percutaneous thrombolytic device. uri icon

Overview

abstract

  • A rubber tip disconnection of Arrow-Trerotola percutaneous thrombolytic device (PTD) may occur occasionally. We experienced 5 cases of a rubber tip disconnection among 453 mechanical thrombectomy sessions with the use of PTD. We present a report about these five cases and suggest possible causes for the occurrences.

publication date

  • March 7, 2014

Research

keywords

  • Catheter Obstruction
  • Equipment Failure Analysis
  • Renal Dialysis
  • Thrombectomy