Lack of early dislocation following total hip arthroplasty with a new dual mobility acetabular design. Academic Article uri icon

Overview

abstract

  • Dual mobility implant designs minimise the risk of dislocation without sacrificing range of motion. Between 1st September 2008 and 31st July 2011, 5 institutions examined early clinical outcomes of a new dual mobility bearing hip system implanted in 485 primary THAs in 452 patients. Patient demographics were 46% female, a mean age of 67 years and a mean BMI of 30. Complications at a minimum of 2 years after surgery included 1 femur fracture, 1 DVT and 4 unrelated deaths. There were no dislocations. For functional outcomes, Harris Hip Scores increased from 41 to 86 (p<0.001), while VAS pain scores decreased from 5.9 to 0.7 (p<0.001). Minimal complications, excellent early clinical outcomes and the absence of early dislocations demonstrate the improved stability of this dual mobility implant system.

publication date

  • February 3, 2015

Research

keywords

  • Arthroplasty, Replacement, Hip
  • Hip Dislocation
  • Hip Prosthesis
  • Osteoarthritis, Hip
  • Postoperative Complications
  • Range of Motion, Articular

Identity

Scopus Document Identifier

  • 84928138422

Digital Object Identifier (DOI)

  • 10.5301/hipint.5000186

PubMed ID

  • 25655740

Additional Document Info

volume

  • 25

issue

  • 1