A snare in time--adding a safety mechanism to the Amplatzer Septal Occluder. Editorial Article uri icon

Overview

abstract

  • Delivery cable/attachment of the Amplatzer Septal Occluder holds the device at a suboptimal angle relative to the plane of the atrial septum prior to release A small gooseneck snare tightened on the hub can maintain flexible control of the device after release of the delivery cable This technique can improve safety of delivery in difficult cases if you think ahead.

publication date

  • April 1, 2016

Research

keywords

  • Heart Septal Defects, Atrial
  • Septal Occluder Device

Identity

Scopus Document Identifier

  • 84963819659

Digital Object Identifier (DOI)

  • 10.1002/ccd.26533

PubMed ID

  • 27085031

Additional Document Info

volume

  • 87

issue

  • 5