Dose parameters of 125I and 192Ir seed sources. Academic Article uri icon

Overview

abstract

  • As mandated by an NCI brachytherapy contract, we measured dosimetric parameters for 192Ir seeds and two models of 125I seeds. Measurements were with LiF powder in a water-equivalent phantom. Data were corrected for background, sample mass, and finite detector volume. Selected parameters were also investigated through Monte Carlo calculations. Results are presented in terms of a dose parametrization that is described in detail, and are compared to published data. Our results agreed well with published data for relative quantities such as radial and angular dose dependence. Our measured value for the 192Ir dose factor was 4.55 cGy(H2O) cm2 mCi-1 h-1, also in good agreement with commonly used values. However, the measured dose factors for 125I seed models 6702 and 6711 were 1.18 and 1.06 cGy(H2O) cm2 mCi-1 h-1, values well below those in general use.

publication date

  • January 1, 1989

Research

keywords

  • Brachytherapy
  • Iodine Radioisotopes
  • Iridium Radioisotopes

Identity

Scopus Document Identifier

  • 0024473383

Digital Object Identifier (DOI)

  • 10.1118/1.596322

PubMed ID

  • 2770636

Additional Document Info

volume

  • 16

issue

  • 4