Long-term failure of Amplatzer plugs in the treatment of aortic pathology. Academic Article uri icon

Overview

abstract

  • We present two cases of Amplatzer® device deployment that exhibited long-term failure necessitating surgical correction. The first case illustrates recanalization of a ruptured Valsalva aneurysm with worsening aortic insufficiency after percutaneous repair with the Amplatzer Vascular Plugs and Amplatzer duct occluder, while the second case illustrates recanalization after deployment of the Amplatzer septal occluder in a patient with a pseudoaneursym in a bovine aorta after type I aortic dissection repair. These two cases illustrate the potential long-term complications and limitations in durability with these devices.

publication date

  • June 23, 2017

Research

keywords

  • Aneurysm, False
  • Aorta
  • Aortic Rupture
  • Aortic Valve Insufficiency
  • Equipment Failure
  • Septal Occluder Device
  • Sinus of Valsalva