Standardized Definition of Structural Valve Degeneration for Surgical and Transcatheter Bioprosthetic Aortic Valves. Review uri icon

Overview

abstract

  • Bioprostheses are prone to structural valve degeneration, resulting in limited long-term durability. A significant challenge when comparing the durability of different types of bioprostheses is the lack of a standardized terminology for the definition of a degenerated valve. This issue becomes especially important when we try to compare the degeneration rate of surgically inserted and transcatheter bioprosthetic valves. This document, by the VIVID (Valve-in-Valve International Data), proposes practical and standardized definitions of valve degeneration and provides recommendations for the timing of clinical and imaging follow-up assessments accordingly. Its goal is to improve the quality of research and clinical care for patients with deteriorated bioprostheses by providing objective and strict criteria that can be utilized in future clinical trials. We hope that the adoption of these criteria by both the cardiological and surgical communities will lead to improved comparability and interpretation of durability analyses.

authors

publication date

  • January 23, 2018

Research

keywords

  • Bioprosthesis
  • Heart Valve Diseases
  • Heart Valve Prosthesis
  • Heart Valve Prosthesis Implantation
  • Heart Valves
  • Prosthesis Failure
  • Terminology as Topic
  • Transcatheter Aortic Valve Replacement

Identity

Scopus Document Identifier

  • 85045031054

Digital Object Identifier (DOI)

  • 10.1161/CIRCULATIONAHA.117.030729

PubMed ID

  • 29358344

Additional Document Info

volume

  • 137

issue

  • 4