Indirect calorimetry in the mechanically ventilated patient.
Academic Article
Overview
abstract
We used indirect calorimetry to measure oxygen consumption (VO2) and carbon dioxide production in 29 mechanically ventilated patients. These data were compared to VO2 measured simultaneously by a standard thermodilution technique. A good correlation was demonstrated between the methods, but VO2 measured by indirect calorimetry was 15% higher than VO2 measured by thermodilution.