The effect of energy window on cardiac ejection fraction. Academic Article uri icon

Overview

abstract

  • ECG gated gamma-ray energy spectra from the left ventricle were created each 50 msec during the cardiac cycle. Nine of ten subjects were studied with a nonimaging Nal probe, and the tenth with a high-resolution Germanium detector. Placing multiple energy windows over the energy spectra, EF was found to vary with the energy window selected. Moving a 20% window across the photopeak produced a roughly linear increase in EF with energy (2.3 EF units per 10 keV increase in energy) in eight of the ten subjects. Dividing the photopeak into a low (126-140 keV) and high-energy (140-154 keV) portion gave significantly different EFs (high energy exceeding low energy by 17%). Increasing the width of a narrow window centered about the photopeak produced negligible change in EF. Examining the energy spectra showed that the small-angle scattered radiation (126-139 keV) was proportionately greater at end systole than at end diastole, after normalizing the spectra to the same photopeak area.

publication date

  • March 1, 1988

Research

keywords

  • Heart
  • Stroke Volume