A rare case of persistent pseudohypobicarbonatemia arising from chemistry analyzer-specific interference. uri icon

Overview

abstract

  • BACKGROUND: Major discrepancies between concentrations of serum total carbon dioxide (tCO2) obtained from chemistry analyzers and calculated bicarbonate from blood gas analyzers should prompt laboratory investigation. Here, we present a rare case of pseudohypobicarbonatemia unrelated to the common causes such as hypertriglyceridemia and hyperproteinemia, but was caused by a low concentration of paraproteins. CASE: A 75-year-old man with persistent fevers was found to have a low concentration of serum tCO2 (<10 mmol/l) with a normal pH and calculated bicarbonate concentrations (23.5 mmol/l) from the blood gas analyzer. His serum tCO2 concentrations remained critically low throughout hospitalization without any evidence of acidosis. Serum tCO2 levels were measured via Siemens ADVIA Chemistry XPT system. RESULTS: Mixing studies revealed non-linearity of serum tCO2, suggesting the presence of interfering substances. Triglyceride concentrations were normal. Serum electrophoresis revealed a 0.4 mg/dl M-protein. The patient's serum tCO2 concentrations were repeated on different chemistry analyzer platforms - including Siemens, Roche, and Abbott - which demonstrated that the interference was specific to the Siemens chemistry analyzer. Serum tCO2 was significantly elevated after ultrafiltration of paraprotein, which confirmed the root cause of pseudohypobicarbonatemia. CONCLUSION: Laboratory professionals should be aware that spuriously low serum tCO2 concentrations may result from unique interfering substances, such as paraproteins, that are both patient- and chemistry analyzer-specific.

publication date

  • May 26, 2021

Research

keywords

  • Acidosis
  • Carbon Dioxide

Identity

Scopus Document Identifier

  • 85107406598

Digital Object Identifier (DOI)

  • 10.1016/j.cca.2021.05.025

PubMed ID

  • 34051269

Additional Document Info

volume

  • 519