Dual flip-angle IR-FLASH with spin history mapping for B1+ corrected T1 mapping: Application to T1 cardiovascular magnetic resonance multitasking. Academic Article uri icon

Overview

abstract

  • PURPOSE: B METHODS: T RESULTS: For phantom-on-table T1 estimates, 2FA agreed better with IR-TSE (intraclass correlation coefficient [ICC] = 0.996, mean error ± SD = -1.6% ± 1.9%) than did 1FA (ICC = 0.922; mean error ± SD = -4.3% ± 12%). For phantom-on-thorax, 2FA was more repeatable and robust to respiration than 1FA (coefficient of variation [CoV] = 1.2% 2FA, = 11.3% 1FA). In vivo, in intrasession T1 repeatability, 2FA (septal CoV = 2.4%, six-segment CoV = 4.4%) outperformed 1FA (septal CoV = 3.1%, six-segment CoV = 5.5%). In six-segment T1 homogeneity, 2FA (CoV = 7.9%) also outperformed 1FA (CoV = 11.1%). CONCLUSION: B

publication date

  • July 26, 2021

Research

keywords

  • Image Interpretation, Computer-Assisted
  • Magnetic Resonance Imaging

Identity

PubMed Central ID

  • PMC8568626

Scopus Document Identifier

  • 85111377490

Digital Object Identifier (DOI)

  • 10.1186/s12968-016-0302-x

PubMed ID

  • 34309072

Additional Document Info

volume

  • 86

issue

  • 6