A reproducible sodium-lithium countertransport assay: the outcome of changing key laboratory parameters. Academic Article uri icon

Overview

abstract

  • This paper describes experimental conditions for reproducible measurement of sodium-lithium countertransport in red blood cells. The assay is sensitive to temperature (10% per degree C) and the condition of the red cells; it is affected little by changes in intracellular lithium over the range 2-7 mmol/1 or by red cell concentrations with laboratory hematocrits of 0.03 to 0.07. Repeated measurements of the same subjects give day-to-day coefficients of variation of 10% or less. The mean difference for interlaboratory comparisons is 11%.

publication date

  • July 15, 1982

Research

keywords

  • Erythrocytes
  • Lithium
  • Sodium

Identity

Scopus Document Identifier

  • 0019980956

Digital Object Identifier (DOI)

  • 10.1016/0009-8981(82)90135-8

PubMed ID

  • 7105417

Additional Document Info

volume

  • 122

issue

  • 3