Particulate metallic debris in cemented total hip arthroplasty. Review uri icon

Overview

abstract

  • Several studies conducted by the authors in the last six years demonstrate that the generation of metallic debris is more severe with titanium alloy than with cobalt-chrome alloy femoral components in cemented total hip arthroplasty (THA). The debris is generated from the articulating surface, particularly if entrapped acrylic debris produces three-body wear, and from the stem surface when the component loosens and abrades against fragmented cement. In selected cases in which the titanium metallic debris is copious, premature failure and severe progressive bone loss occurs. Electron microscopy demonstrates that the particles of metallic debris can be extremely small (a few hundredths of 1 micron). They are phagocytized by the macrophages and transported to the phagolysosomes. In this highly corrosive environment, the very high surface area of the particles may release toxic concentrations of the constituents of the alloy intracellularly, probably leading to progressive cell degeneration and death, with subsequent release of intracellular enzymes and ingested metallic debris. This cycle most likely repeats itself, leading to tissue necrosis. The results presented do not support the use of titanium alloy femoral components for cemented THA, particularly for the articulating surface.

publication date

  • August 1, 1993

Research

keywords

  • Bone Cements
  • Chromium Alloys
  • Hip Prosthesis
  • Polyethylenes
  • Stainless Steel
  • Titanium

Identity

Scopus Document Identifier

  • 0027296132

PubMed ID

  • 8339477

Additional Document Info

issue

  • 293