Arterial-phase three-dimensional gadolinium magnetic resonance angiography of the renal arteries. Strategies for timing and contrast media injection: original investigation.
Academic Article
Overview
abstract
RATIONALE AND OBJECTIVES: The authors review different imaging and contrast-media infusion strategies for arterial-phase three-dimensional (3D) gadolinium-enhanced magnetic resonance angiography (Gd-MRA). METHODS: The influence of physicochemical factors on the infusion of contrast media, including viscosity, flow rate, inline pressure, and cannula size, is assessed. The combination of manual or automated contrast-media administration with timing-dependent or -independent 3D Gd-MRA techniques is reviewed regarding the aspects of effectiveness, robustness, image quality, and costs. RESULTS: For effective bolus delivery with high flow rates, the type and temperature of the contrast media, the size of the cannula, and an immediate saline flush must be considered. Timing-dependent techniques based on a test bolus and using automated contrast-media infusion as well as timing independent techniques such as MR SmartPrep or multiphase 3D Gd-MRA by using a manual injection with a SmartSet tubing set, are all effective procedures for arterial phase 3D Gd-MRA. CONCLUSIONS: Manual contrast-media injection with a tubing set can be used for timing-independent MRA techniques. The multiphase 3D Gd-MRA approach seems to be favorable for different MR systems, robustness, and speed.